The reported event could be confirmed, although the device was not returned for evaluation, but a device picture and intraoperative x-ray were shared by the customer through which we can confirm the event.In the provided picture, k-wire shows broken and indicated towards oblique insertion and forceful extraction.More detailed information as well as the affected device must be returned to determine the exact root cause.The provided device images and intra-op x-rays without multiple views are not enough to determine the exact root cause.One of the possible reasons of failure could be as provided in the picture, k-wire shows broken and indicated towards oblique insertion and forceful extraction.A review of the device history for the reported lot did not indicate any abnormalities.No indications of manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.If the device is returned or if any additional information is provided, the investigation will be reassessed.Device was discarded.
|