MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97715
Device Problems Failure to Deliver Energy (1211); Charging Problem (2892); Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/14/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer concerning patient with an implantable neurostimulator (ins).Patient stated for the past 3 days she noticed that her ins charge is not depleting.Patient states she normally charged once/day but has not had to charge ins.Patient states for 2 days, both controller and ins battery indicated 100%.During call, the patient was assisted with connecting the controller to ins which indicated stimulation is off.The patient could not confirm how stimulation was turned off.Patient turned stim on and will monitor charge frequency.No patient symptoms were reported.No further complications were reported/anticipated.
 
Event Description
Additional information was received from the patient.It was reported that the device would not charge.There were no changes that led to the issue.The issue was reported to be resolved.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key9772538
MDR Text Key181916311
Report Number3004209178-2020-04367
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2020
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Date Manufacturer Received03/09/2020
Date Device Manufactured05/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient Weight250
-
-