Model Number 9445-02
Device Problem No Device Output (1435)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/14/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that transmitter failed error occurred.Data was evaluated and the allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that transmitter failed error occurred.The product was evaluated.An external visual inspection was performed and passed.Voltage check was performed and passed.Pairing with (b)(4) bluetooth device was performed and passed.A review of the share logs was performed and transmitter failed error was not found within the investigation window.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key9772539
MDR Text Key181532377
Report Number3004753838-2020-020185
Device Sequence Number1
Product Code QBJ
UDI-Device Identifier00386270000385
UDI-Public00386270000385
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170088
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/01/2020
Device Model Number9445-02
Device Catalogue NumberSTT-OE-002
Device Lot Number5262474
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/16/2020
Initial Date FDA Received02/29/2020
Supplement Dates Manufacturer Received06/01/2020
Supplement Dates FDA Received06/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age62 YR
-
-