• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER TORQVUE DELIVERY SYSTEM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER TORQVUE DELIVERY SYSTEM; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 9-ITV06F180/80
Medical Device Problem Codes Mechanical Problem (1384); Retraction Problem (1536)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 01/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
Related manufacturer reference number 2135147-2020-00086.On (b)(6) 2020, an 8mm amplatzer muscular vsd was attempted to be implanted.It was mis-sized too small and exchanged for a 10mm amplatzer muscular vsd occluder (lot # 6925807) with a 6f amplatzer torqvue delivery system.When positioning the device, it was determined to be mi-sized too small.When attempting to retract the device, the device would not re-sheath nor retract into the 6f delivery sheath and a vascular issue occurred with hematoma at the groin.The device and the delivery system was removed and the delivery system was exchanged for an 8f torqvue 45/80 delivery system (lot # unknown) and a new 10mm muscular vsp (lot # 6942246).The physician noted that the patient had a torturous anatomy and there was no kinks noted on the first delivery system (lot # 6271477).There was a clinically significant delay in the procedure.The patient is stable during and post procedure.
 
Additional Manufacturer Narrative
Additional information for : d10, g4, g7, h2, h3, h6, and h10.An event of difficulty retracting the occluder into the sheath was reported.The investigation found that the soft tip of the sheath was damaged and inverted, which is consistent with difficulties retracting the device.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TORQVUE DELIVERY SYSTEM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key9772694
Report Number2135147-2020-00085
Device Sequence Number10956945
Product Code DQY
UDI-Device Identifier00811806010779
UDI-Public00811806010779
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K072313
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model Number9-ITV06F180/80
Device Catalogue Number9-ITV06F180/80
Device Lot Number6271477
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/03/2020
Supplement Date Received by Manufacturer03/28/2020
Initial Report FDA Received Date03/01/2020
Supplement Report FDA Received Date04/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
9-VSD-MUSC-010.
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-