| Model Number |
MMT-751NAH |
| Medical Device Problem Codes |
Mechanical Problem (1384); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump was sudden power down also it lost the settings.Customer also stated that insulin pump was slow to rewind and louder during the rewind.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device monitored for several days.Device received with all operating currents within specification.Device passed a21 error test, rewind test, displacement test.No unexpected battery power loss anomalies noted.No unexpected rewind anomalies noted.(b)(4).
|
| |
|