• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 HYPERGLIDE; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 HYPERGLIDE; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY Back to Search Results
Model Number 104-4112
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since the device was not returned, we are unable to perform further root cause analysis and the exact cause of the reported event is unknown.All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.No corrective action.Monitoring and trending this type of event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received report that the pipeline flex with shield device did not open at the proximal section.The hyperglide balloon was selected to open the pipeline device.However, the balloon ruptured during the inflation.A new balloon was used to open the pipeline device.No patient injury occurred.The vessel anatomy was normal in tortuosity.The pipeline was place in a curb / bend in the anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYPERGLIDE
Common Device Name
CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key9774241
Report Number2029214-2020-00179
Device Sequence Number19379482
Product Code MJN
UDI-Device Identifier00847536004863
UDI-Public00847536004863
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
K021066
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2022
Device Model Number104-4112
Device Lot NumberA848770
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2020
Initial Report FDA Received Date03/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-