• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOUND TREE MEDICAL, LLC CURAPLEX; AIRWAY, OROPHARYNGEAL, ANESTHESIIOLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOUND TREE MEDICAL, LLC CURAPLEX; AIRWAY, OROPHARYNGEAL, ANESTHESIIOLOGY Back to Search Results
Model Number #477KLTSD4125
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/26/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Curaplex supraglotic airway kit, king lts-d, size 2.5, 25-35kg #477-kltsd4125 (lot #asm0021574) contains: (1) supraglotic airway, king lts-d size 2.5 (1) syringe only, 10 oc luer lock (1) gastric tube, 16fr (1) packet of lubricating jelly.The manufacturer (ambu) of the king airway lts-d recommends an inflation volume of 30-40 cc for the device and the packaged product from the manufacturer comes with a syringe sized to support this volume.The 10 cc syringe found in the curaplex kit with the same king airway device could result in an under-inflated balloon if the user does not re-load the syringe with air 3-4 times.An under-inflated balloon could lead to device failure resulting in hypoxia and if untreated death.Curaplex product by samova hc.Llc's family of companies: bound tree medical llc.Cardic partners, inc.Emergency medical products, inc.& tri-anim health services inc.Distributor: (b)(4).Fda safety report id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURAPLEX
Common Device Name
AIRWAY, OROPHARYNGEAL, ANESTHESIIOLOGY
Manufacturer (Section D)
BOUND TREE MEDICAL, LLC
MDR Report Key9774578
Report NumberMW5093396
Device Sequence Number11376651
Product Code CAE
UDI-Device Identifier00817617020549
UDI-Public(01)00817617020549
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Voluntary
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model Number#477KLTSD4125
Device Catalogue Number477-KLTSD4125
Device Lot NumberASM0021574
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability;
-
-