Model Number 1960-01-400
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Codes Not Applicable (3189); No Code Available (3191)
Date of Event 02/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
When opening the implant box the implant was covered with fluff.A new implant was opened.The surgery was delayed for about 10 minutes due to the reported event.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> examination of the returned device confirmed the reported event.Root cause and corrective action are documented in the capa system.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Root cause and corrective action are documented in the capa system.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIG CR NPOR FEM RT SZ 4N
Common Device Name
SIGMA KNEE PRIMARY : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9776059
Report Number1818910-2020-06532
Device Sequence Number2330812
Product Code JWH
UDI-Device Identifier10603295074779
UDI-Public10603295074779
Combination Product (Y/N)N
Initial Reporter CountrySE
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1960-01-400
Device Catalogue Number196001400
Device Lot NumberJ4723X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2020
Initial Date Received by Manufacturer 02/17/2020
Supplement Date Received by Manufacturer03/19/2020
04/06/2020
Initial Report FDA Received Date03/02/2020
Supplement Report FDA Received Date03/27/2020
04/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
Patient Weight60
-
-