| Model Number |
1294-31-125 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Pain (1994); No Code Available (3191)
|
| Date of Event |
02/14/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Revision right tkr.Patient underwent right pfc rp, total knee replacement on (b)(6) 2016.Presented to surgeon with complaint of tibial pain and instability.Suspected loose tibial tray.Upon opening the knee, surgeon stated it appeared the tray had subsided on the lateral side.Surgeon proceeded to revise the tibial tray component only to mbt tray with tibial sleeve and stem and upsized the insert.No ae reported, no delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. h10 additional narrative: udi: (b)(4).Added: d4(expiration date).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|