| Model Number |
AIRLIFE ADULT MANUAL RESUSCITATOR, 40'' (1.0M) OXYGEN RESERVOIR TUBING, ADULT M |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Hypoxia (1918)
|
| Date of Event |
01/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As of this time, no sample has been returned.There are three(3) lot numbers available for this sample: 0004118094, 0004111662 and 0004095580.Any additional information received from the customer will be included in a follow-up report.
|
| |
|
Event or Problem Description
|
|
The customer reported that the airlife adult manual resuscitator experienced a defect problem.In post anesthesia care unit, patient experienced difficulty breathing.So they bagged the patient to continue breathing.But due to hose at base of ambu bag was kinked or pinched, it did not allow adequate oxygen to flow into the ambu bag.When the issue occurred, the patient's spo2 were not coming up from low 80%, once identified staff were able to get line open enough to bring patient's spo2 up to 99%.
|
| |
|