• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE VISIONPRO 26" LED DISPLAY INTERNATIONAL KIT; CAMERA, TELEVISION, SURGICAL, WITHOUT AUDIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE VISIONPRO 26" LED DISPLAY INTERNATIONAL KIT; CAMERA, TELEVISION, SURGICAL, WITHOUT AUDIO Back to Search Results
Model Number 0240031020
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 01/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was image loss during procedure.Please note that the procedure was completed successfully.
 
Event or Problem Description
It was reported that there was image loss during procedure.Please note that the procedure was completed successfully.
 
Additional Manufacturer Narrative
The device was received at the local service facility for investigation.The visionpro display under test power up initially showing a normal image but then the image turns black while the power led and backlight remains on.If left on, the image returns to normal and becomes black again at random intervals.The faulty visionpro display will be discarded and replaced with a new one.Faulty mode - faulty display faulty cause - faulty monitor probable root cause: display main board / sub board, connectors/ cables, display firmware, display power supply, display lcd panel, use error, electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge, manufacturing/ service nonconformity.The reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONPRO 26" LED DISPLAY INTERNATIONAL KIT
Common Device Name
CAMERA, TELEVISION, SURGICAL, WITHOUT AUDIO
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key9777483
Report Number0002936485-2020-00102
Device Sequence Number3267916
Product Code FWB
UDI-Device Identifier07613327056082
UDI-Public07613327056082
Combination Product (Y/N)N
Initial Reporter CountryEG
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0240031020
Device Catalogue Number0240031020I
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2020
Initial Date Received by Manufacturer 02/04/2020
Supplement Date Received by Manufacturer02/04/2020
Initial Report FDA Received Date03/02/2020
Supplement Report FDA Received Date04/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-