The device was received at the local service facility for investigation.The technical service report indicates: the reported event is confirmed.The visionpro display under test power up initially showing a normal image but then the image turns black while the power led and backlight remains on.If left on, the image returns to normal and becomes black again at random intervals.The faulty visionpro display will be discarded and replaced with a new one.Faulty mode - faulty display faulty cause - faulty monitor probable root cause: display main board / sub board, connectors/ cables, display firmware, display power supply, display lcd panel, use error, electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge, manufacturing/ service nonconformity.The reported failure mode will be monitored for future reoccurrence.
|