| Model Number |
34310MS-D |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device has not yet been evaluated; it is in transit to the manufacturer, (b)(4).
|
| |
|
Event or Problem Description
|
|
As per a vigilance report filed with national competent authority, swissmedic, by the factory in (b)(6): during a laparoscopic procedure, the jaws of the clickline scissors insert could no longer be closed intra-abdominally.To remove the instrument, the trocar had to be removed and a pair of pliers was used to close the jaws in order to bring the scissors insert through the orifice.
|
| |
|
Additional Manufacturer Narrative
|
|
As per the manufacturer's evaluation: upon evaluation the scissor blades were found to be damaged.A hardness test was unobtrusive.The scissor blades are dented in a way they cannot be closed anymore.The damages of the two cutting edges overlap exactly.Most probably the scissor blades were cutting themselves due to high force initiation and bending of the blades.The device was manufactured in april 2019.The root cause most probably is too high force initiation and bending of the blades.No indication for a material or manufacturing related issue was found during the investigation.
|
| |
|