• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR WORLDWIDE LLC MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR WORLDWIDE LLC MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TMPX215
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); No Code Available (3191)
Date of Event 01/31/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient who underwent a breast augmentation revision procedure with a mentor memorygel xtra breast implant 215cc gel breast prosthesis developed baker grade iii capsular contracture in her left breast.As a result, the patient underwent explantation and replacement with a mentor memorygel breast implant 190cc gel breast prosthesis on (b)(6) 2020.
 
Additional Manufacturer Narrative
On (b)(6) 2020, the mentor failure analysis lab received the device for evaluation.On (b)(6) 2020, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information provided, the patient experienced capsular contracture on the left breast.During the visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Mentor concluded that the capsular contracture in the patient¿s breast was the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of our quality process, the manufacturing records of this lot-serial number were reviewed and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Capsular contracture complaint information is consistently analyzed and monitored by quality assurance to determine when further action is necessary.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL XTRA BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR WORLDWIDE LLC
33 technology drive
irvine 92618
MDR Report Key9777775
Report Number3006942524-2020-00002
Device Sequence Number19529970
Product Code FTR
UDI-Device Identifier00081317025429
UDI-Public00081317025429
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/21/2023
Device Model NumberTMPX215
Device Catalogue NumberTMPX215
Device Lot Number7540711
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2020
Initial Date Received by Manufacturer 02/10/2020
Supplement Date Received by Manufacturer10/20/2020
Initial Report FDA Received Date03/02/2020
Supplement Report FDA Received Date11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
-
-