| Model Number |
MMT-751NAL |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the up buttons were unresponsive.The customer also stated that the battery compartment was cracked.Troubleshooting was performed for damage.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with broken battery tube thread noted.The test reservoir was able to lock in place.Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device passed displacement test and self test.
|
| |
|