• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 3 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 3 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-04-030
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cyst(s) (1800); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Unspecified Infection (1930); Pain (1994); Osteolysis (2377); No Code Available (3191)
Date of Event 04/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pcf and medical records.After review of medical records patient was revised to addressed infected right total hip arthroplasty, metallosis and adverse local tissue reaction.Operative notes indicated severe purulent in intraarticular area as well as the region between the adductor tendons and fascia lata.A pocket of pus anterior to the acetabulum.There was extensive osteolysis involving both acetabulum and proximal femur.There is no erosion in adductor tendons as inserted to the greater trochanter.Trunnionosis with black material about the head and neck junction.A fatty large lytic lesion on the medial neck of femur was noted.Debridement of pseudotumor and soft tissues were performed as well as the necrotic bone.Doi: (b)(6) 2010; dor: (b)(6) 2018 right hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot null device history batch null device history review null if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 3 STD OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9779452
Report Number1818910-2020-06614
Device Sequence Number2650421
Product Code KWA
UDI-Device Identifier10603295000747
UDI-Public10603295000747
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K073570
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2019
Device Model Number1012-04-030
Device Catalogue Number101204030
Device Lot NumberD65BL1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/14/2020
Supplement Date Received by Manufacturer04/06/2020
Initial Report FDA Received Date03/03/2020
Supplement Report FDA Received Date04/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
-
-