Medical Device Problem Codes Difficult to Insert (1316); Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: (b)(6).Complaint sample was evaluated and the reported event was unable to confirmed.The device was not returned, however a thread like material which appears to be the barb that is used to lock the liner into the ring was returned.Dhr was reviewed and no discrepancies were found.Root cause was unable to be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during a hip procedure, a liner was driven into a shell and was unable to be secured.Once the liner was taken out, a thread-like material came out allowing the liner to seat within the shell.There was no reported harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER 32MM F
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9779589
Report Number0001825034-2020-00779
Device Sequence Number504305
Product Code PBI
UDI-Device Identifier00880304527041
UDI-Public00880304527041
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/11/2022
Device Model NumberN/A
Device Catalogue Number010000928
Device Lot Number6167741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2020
Initial Report FDA Received Date03/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-