• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. 3D KNEE; INSERT, 3D EX SZ 6LT 9MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. 3D KNEE; INSERT, 3D EX SZ 6LT 9MM Back to Search Results
Model Number 391-09-706
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - the patient's knee became unstable and a thicker insert needed to be inserted.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as instability.The previous surgery and the surgery detailed in this event occurred 2 months and 2 weeks apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr# (b)(4) associated with the main part #391-09-706, insert, 3d ex sz 6lt 9mm, which documents that out of 20 parts lot, 1 part was rejected and scrapped as the part had an inclusion that can be felt using shim fg587.All other items in the lot were met with the design, fit and function requirements.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to instability.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3D KNEE
Common Device Name
INSERT, 3D EX SZ 6LT 9MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key9779827
Report Number1644408-2020-00150
Device Sequence Number10941170
Product Code OIY
UDI-Device Identifier00888912118965
UDI-Public(01)00888912118965
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K091956
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/21/2024
Device Model Number391-09-706
Device Catalogue Number391-09-706
Device Lot Number283G1985
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/04/2020
Supplement Date Received by Manufacturer03/05/2020
Initial Report FDA Received Date03/03/2020
Supplement Report FDA Received Date04/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight95
-
-