• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 11/13 S-ROM 40MM M SPEC+9; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 11/13 S-ROM 40MM M SPEC+9; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-45-500
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Codes Adhesion(s) (1695); Edema (1820); Foreign Body Reaction (1868); Pain (1994); No Code Available (3191)
Date of Event 05/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Initial reporter occupation: lawyer.
 
Event or Problem Description
Pinnacle claim submission form and medical records received.After review of medical records, the patient was revised to address a failed right metal on metal total hip arthroplasty with alval/ fluid collection noted on mri and pain.Preop workup for infection was negative.Intraoperatively, there was about 20cc of yellow fluid that was suctioned along with grey/ metal stained reactive tissue present.There was no sign of loosening on the stem and shell and the trunnion had no obvious damage/ fracture/ wear.There was extensive scar tissue around the cup that was debrided.There was no evidence of infection.There was no black staining inside the metal head and on the back of the liner.Doi: (b)(6) 2010.Dor: (b)(6) 2018, (right hip).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint#: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11/13 S-ROM 40MM M SPEC+9
Common Device Name
S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9780367
Report Number1818910-2020-06679
Device Sequence Number2788062
Product Code JDI
UDI-Device Identifier10603295033851
UDI-Public10603295033851
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1365-45-500
Device Catalogue Number136545500
Device Lot Number2471718
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/13/2020
Supplement Date Received by Manufacturer04/02/2020
Initial Report FDA Received Date03/03/2020
Supplement Report FDA Received Date04/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
-
-