• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3562
Device Problems High impedance (1291); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/12/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that during a device upgrade, the left ventricular lead exhibited high, out of range impedance.Upon investigation, the physician determined the set screw connection was the problem and elected to use a different pacemaker.The new device was successfully implanted.The patient was stable before, during and after the procedure.
 
Manufacturer Narrative
The reported field event of high impedance was not confirmed in the laboratory.The device was tested on the bench and using automated testing equipment, and no anomalies were found.The lv-setscrew was found normal and tightened onto the lv test leads normally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key9780718
MDR Text Key181930910
Report Number2017865-2020-02208
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000093910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/12/2020
Initial Date FDA Received03/03/2020
Supplement Dates Manufacturer Received06/23/2020
Supplement Dates FDA Received06/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Weight82
-
-