Model Number V60
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 03mar2020.
 
Event Description
It was reported that the ventilator experienced a 35 volt failure.There was no patient involvement.The field service engineer (fse) evaluated the ventilator.The fse replaced the power management printed circuit board assembly (pm pcba) and the problem was resolved.
 
Manufacturer Narrative
G4: 28mar2020.B4: 31mar2020.The replaced power management printed circuit board assembly (pm pcba) was returned for failure analysis.The pm pcba was tested and no failures were identified.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9780720
MDR Text Key194059277
Report Number2031642-2020-00713
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-