| Model Number |
401932 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to stryker sustainability solutions for evaluation.Upon visual inspection of the received complaint device, a kink was identified 4.5cm from the distal tip which is distal to the transition point.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The most likely root cause is mishandling subsequent to distribution, including shipping/storage conditions or improper manipulation of the catheter.The instructions for use (ifu) state: do not introduce the tip folded into the guiding sheath.Do not exert excessive pressure during placement of catheter if unknown resistance is encountered.Do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.Avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.The reported event will continue to be monitored through post-market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported that the electrophysiology catheter bent when the physician attempted to deflect it.Additional information provided by the customer revealed the electrophysiology catheter had a kink amidst the electrodes.There was no patient injury or medical intervention and the extended procedure time reported was less than 30 minutes to replace the device.These are commonly used devices that are readily available.
|
| |
|