| Model Number |
357464 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A passeo-18 balloon catheter was chosen for treatment.During removal of the protection cap resistance was felt but it could be finally taken off.After insertion of the device it was not possible to inflate the balloon.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinal over a length of about 29 mm.Microscopic inspection of the balloon showed several deep scratches in close vicinity of the tear.It seems likely that the tear in the balloon was caused by a hard, sharp edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.The instrument was delivered in a leak-proof condition.Based on the conducted investigations no manufacturing related root cause could be determined.
|
| |
|