| Model Number |
366996 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A pantera leo balloon catheter was chosen for treatment of a lesion in the rca.During inflation the balloon ruptured at 14 atm.Therefore, another balloon was used to finish the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.The instrument was delivered in a leak-proof condition.Based on the conducted investigations of the device being subject to this complaint, no manufacturing or material related root cause could be determined.
|
| |
|
Search Alerts/Recalls
|