PLEXUS MANUFACTURING SDN. BHD INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97755
Medical Device Problem Codes Overheating of Device (1437); Delayed Charge Time (2586); Charging Problem (2892)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 02/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins).It was reported that the recharger (rtm) won¿t recognize the ins and a ¿no device found¿ message was displayed.The patient reported that there was difficulty charging the ins and it took 3 hours to charge to about 80%.It was reported that the first inch of the rtm cord that goes into the paddle gets so hot that it burnt her skin.
 
Additional Manufacturer Narrative
H3.Analysis of the recharger (s/n (b)(6)) found a recharge antenna failure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key9781556
Report Number3004593495-2020-00221
Device Sequence Number2649832
Product Code LGW
UDI-Device Identifier00613994925985
UDI-Public00613994925985
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number97755
Device Catalogue Number97755
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2020
Supplement Date Received by Manufacturer05/14/2020
Initial Report FDA Received Date03/03/2020
Supplement Report FDA Received Date05/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-