| Model Number |
2218-36-156 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
02/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Usfda mdr determination:no patient consequences, no serious injury to report.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
A recurring issue with difficult cup position cases issues screwing liners into the acetabular cup.When using the screwdriver, the liner central screw mechanism angles and eventually disassociates the locking ring from the trial.Surgical delay of 5 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot null device history batch null device history review null if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|