Model Number CMCV-010-XLG
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Sepsis (2067)
Date of Event 02/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing review of the device history record for the reported lot shows that all units were quality released on 9/9/2019 having met all internal qc acceptance requirements.All sterilization records, and bioburden testing indicate successful sterilization process, and passing lal and product sterility results allowed the lot to be released having met all criteria for manufacturing release.There were no non-conformances associated with the manufacturing lot during production, sterilization and final packaging.In lieu of a request for the oem supplier dhr review, it is noted that aziyo processes the non-sterile envelope materials by cutting, suturing, packaging, and sterilizing.It can be noted that per the instructions for use [ifu] (part number-20663 provided with device) for the cangaroo envelope currently lists infection as a potential complication associated with the procedure and device.Although the exact cause of the reported event cannot be conclusively determined, infection is a known complication with the cangaroo envelope.
 
Event or Problem Description
Patient had pulse generator and leads explanted due to infection/sepsis.Aziyo product remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANGAROO ENVELOPE
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
AZIYO BIOLOGICS, INC.
1100 old ellis road
ste 1200
roswell, ga
MDR Report Key9781984
Report Number3005619880-2020-00022
Device Sequence Number19495663
Product Code FTM
UDI-Device Identifier10859389005079
UDI-Public10859389005079
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K182255
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2022
Device Model NumberCMCV-010-XLG
Device Lot NumberM19K1385002
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2020
Initial Report FDA Received Date03/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
-
-