Model Number MC2-5070S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product has been returned for evaluation as it remains in-situ.Radiographs provided confirmed the alleged event.No root cause can be confirmed as the device remains in-situ.
 
Event or Problem Description
Information was received that the end cap dislodged from the rod.As per reporter physician decided to continue with distractions.No revision procedure is planned at this time.
 
Additional Manufacturer Narrative
Even though no product has been returned, fsca investigation was performed.And variation in the torque applied to the threaded cap, during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although, all operators followed the assembly procedure, and the torque wrench, indicated 40 in-lbs was applied.The manner of using the torque wrench, resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap.Thereby, creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
 
Additional Manufacturer Narrative
The product was returned and the reported failure mode was confirmed.Visual inspection of the returned product confirmed the end cap was unthreaded from the housing tube.The returned rod was observed to be partially distracted.X-ray imaging and internal component inspection were not applicable.Review of the device history records indicated the rod was functionally tested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise
suite 100
aliso viejo CA 92656
MDR Report Key9783277
Report Number3006179046-2020-00126
Device Sequence Number1866314
Product Code PGN
UDI-Device Identifier00887517834027
UDI-Public887517834027
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Notification,Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5070S
Device Lot Number8041114AAA
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/11/2020
Supplement Date Received by Manufacturer02/11/2020
01/15/2021
Initial Report FDA Received Date03/03/2020
Supplement Report FDA Received Date01/25/2021
02/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1898-2020
Patient Sequence Number1
Patient Age6 YR
-
-