| Model Number |
MC2-5070S |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for evaluation as it remains in-situ.Radiographs provided confirmed the alleged event.No root cause can be confirmed as the device remains in-situ.
|
| |
|
Event or Problem Description
|
|
Information was received that the end cap dislodged from the rod.As per reporter physician decided to continue with distractions.No revision procedure is planned at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Even though no product has been returned, fsca investigation was performed.And variation in the torque applied to the threaded cap, during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although, all operators followed the assembly procedure, and the torque wrench, indicated 40 in-lbs was applied.The manner of using the torque wrench, resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap.Thereby, creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned and the reported failure mode was confirmed.Visual inspection of the returned product confirmed the end cap was unthreaded from the housing tube.The returned rod was observed to be partially distracted.X-ray imaging and internal component inspection were not applicable.Review of the device history records indicated the rod was functionally tested.
|
| |
|