| Model Number |
M00560160 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an injection gold probe was used during a colonoscopy procedure performed on (b)(6) 2020.According to the complainant, during the procedure, they advanced a gold probe down the scope.They tried to activate the probe but no energy was getting to the tip of the probe.They removed the probe and used another one, of a different lot number.While advancing down the scope, and the tip of the probe fell off inside the scope.They retrieve the detached piece from the scope using biopsy forceps.The procedure was completed using a third injection gold probe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
| |
|