Model Number REACT-71
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Infarction, Cerebral (1771); Device Embedded In Tissue or Plaque (3165)
Date of Event 02/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Refer to manufacturer report # 2029214-2020-00190 for details pertaining to the related reportable event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received report that the tip of the react catheter separated and remains in the patient's right m1.This event occurred during a mechanical thrombectomy case.The patient vessel remained open.Two react catheters were used and it is unknown which of these two device experienced the device separation.A851300 and a824582 were two devices that were used in the is case.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received indicated there were attempts made by the physician to remove the broken segment of catheter, but it was unsuccessful.The catheter was pretty well locked in place and so no additional intervention was performed to secure it in place.The patient did experience a slight stroke, but it remained uncertain how related it was to the device or an underlying condition.It was noted the patient was doing well now.
 
Additional Manufacturer Narrative
D10: device available for evaluation - additional information.G4.Date mfr rec ¿ additional information.H2: type of follow up - device evaluation.H3: device evaluated by manufacturer- additional information.H6: codes updated the device was returned for evaluation and the clinical observation was confirmed.The react catheter total length and the useable length were measured and found within specification.Upon visual inspection, no damages or irregularities were found with the react hub.Blood was present within the hub.The react-71 catheter body was found to be flattened from distal end.In addition, the catheter body was found damaged from distal end.The distal marker band/tip was found missing/ separated.It was reported that the distal marker band/tip remains within the patient.The react-71 catheter was flushed, water exited from the distal end.No other anomalies were observed.Based on the device analysis and reported information, we were unable to determined the cause of the event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATHETER REACT-71
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key9785458
Report Number2029214-2020-00191
Device Sequence Number19405359
Product Code DQY
UDI-Device Identifier00847536031784
UDI-Public00847536031784
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K182097
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model NumberREACT-71
Device Catalogue NumberREACT-71
Device Lot NumberA824582
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2020
Initial Date Received by Manufacturer 02/28/2020
Supplement Date Received by Manufacturer03/06/2020
04/19/2020
Initial Report FDA Received Date03/04/2020
Supplement Report FDA Received Date03/08/2020
04/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-