| Model Number |
23AJ-501 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Mitral Insufficiency (1963)
|
| Date of Event |
02/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further information regarding this event has been requested.If returned, investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020 a sjm masters series mechanical heart valve was chosen for the procedure.The physician reported during the procedure, the device was implanted and the user noticed the leaflet detached.Additional information has been requested.2/24 pm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information h10, h3 and h6 ; the reported event of the leaflets fractured and dislodged during rotation was confirmed.One leaflet had dislodged and fractured into multiple pieces, part of which was not returned.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.There was no evidence of material defect in the carbon coating that may have caused or contributed to the fractured leaflet.The cause of the reported event could not be conclusively determined, however it is consistent with damage caused by some external force applied to the leaflet and orifice, which overstressed the carbon material.
|
| |
|