According to the reporter, following an open paracolostomy hernia, the patient had to return to the hospital due to enterocutaneous fistula.This resulted to tissue damage and unanticipated tissue loss.Re-operation, fistula treatment and ileostomy were made to compete the case.The patient still hospitalized.
|
|
H3 evaluation summary: a review of the device history record has been performed and no failure that may relate to the reported conditions have been noted.Especially the record related to the sterilization was found within specifications.The visual examination of the 2 provided pictures show: a fistula on the abdominal wall with an abscess.The skin was red and seems irritated around the fistula.A portion of mesh was visible at the fistula location.This mesh did not have the same textile.The reported condition was confirmed.Additionally, a review of the routine product bioburden and endotoxin monitoring results from (b)(6) 2015 until (b)(6) 2016 has been performed.This review confirmed that the bioburden and endotoxin results were within specifications limits for the concerned period.It should be noted that the mesh was placed in an open approach.The mesh was implanted the (b)(6) 2017.The patient was admitted for complications on (b)(6) 2020 and the complications were enterocutaneous fistula.The reported infection, hematoma and fistula are known potential complications of this type of surgery; this is a procedural related complication that is not normally attributed to any product/process deficiencies.The product ifu which accompanies each device states in chapter ¿possible complications¿ that ¿ the possible complications associated with the use of symbotex¿ composite mesh are those typically associated with surgically implantable mesh: seroma, hematoma, recurrence, adhesions, fistula formation, infection, inflammation, chronic pain, and/or allergic reactions to the components of the product.Potential events associated with state of the art mesh with similar indication may include: organ injury (including bowel and visceral injury), trocar-site herniation, bowel obstruction and urinary retention (related with the use of anesthetics).¿ based on our investigation and a complaint history review, the manufacture of the device is not suspected.There is no indication that there is a defective lot or that this event represents an emerging adverse quality trend.The root cause could not be determined.If additional information is obtained, or the sample is returned, we will re-open this investigation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|