• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION SABER 8MM4CM 90; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION SABER 8MM4CM 90; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 48008004S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint conclusion: an 8mm x 4cm 90cm saber percutaneous transluminal angioplasty (pta) balloon catheter ruptured.There was no reported patient injury.The saber balloon was used for a shunt pta case.The target lesion was the anastomosis part.The lesion had moderate calcification and moderate vessel tortuosity with fifty percent (50%) stenosis.The device was not used for a chronic total occlusion (total occlusion >3 months).The device was replaced with a non-cordis balloon catheter and the procedure was completed.The device was stored and handled as per the instruction for use (ifu).There were no damages noted to the packaging of the device prior to use.There was no difficulty removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product into the patient.The device was prepped normally (i.E.Maintain negative pressure) according to the ifu.The same indeflator was used successfully with other devices.There was no resistance/friction while inserting the balloon through the rotating hemostatic valve or through the guide catheter.There was no difficulty advancing the balloon catheter through the vessel or crossing the lesion.The catheter was able to advance through an acute part and there were no kinks or bends noted on the device.The device was removed intact (in one piece) from the patient¿s body.Additional procedural details were requested but are unknown.The product was not returned for analysis as it was discarded due to suspicious infectious disease.A product history record (phr) review of lot 17708885 revealed no anomalies or non-conformance's during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.It is likely vessel characteristics, although unknown may have contributed to the reported event.The procedure performed was a shunt percutaneous transluminal angioplasty (pta) case.Arteriovenous shunts are often scarred and fibrous in nature and are therefore often resistant to balloon expansion increasing the likelihood of damage to the balloon.However, without the return of the device for analysis, it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Event or Problem Description
As reported, an 8mm x 4cm 90cm saber percutaneous transluminal angioplasty (pta) balloon catheter ruptured.Therefore, the device was removed intact (in one piece) from the patient¿s body and replaced with a non-cordis balloon catheter and the procedure was completed.There was no reported patient injury.The saber balloon was used for a shunt pta case.The device was stored and handled as per the instruction for use (ifu).There were no damages noted to the packaging of the device prior to use.There was no difficulty removing the product from the hoop.There was no difficulty removing the protective balloon cover.There was no difficulty removing the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The device was prepped according to the ifu.The device was prepped normally (i.E.Maintain negative pressure).The same indeflator was used successfully with other devices.There was no resistance/friction while inserting the balloon through the rotating hemostatic valve.There was no resistance/friction while inserting the balloon through the guide catheter.The target lesion was the anastomosis part.The lesion had moderate calcification.There was moderate vessel tortuosity with fifty percent (50%) stenosis.The device was not used for a chronic total occlusion (total occlusion >3 months).There was no difficulty advancing the balloon catheter through the vessel.There was no difficulty crossing the lesion.The catheter was able to advance through an acute part and there was no kinked nor bends noted on the device.The device will not be returned for evaluation as it was discarded due to suspicious infectious disease.Additional procedural details were requested but are unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER 8MM4CM 90
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
MDR Report Key9786295
Report Number9616099-2020-03549
Device Sequence Number687778
Product Code LIT
UDI-Device Identifier20705032068908
UDI-Public20705032068908
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K971010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2020
Device Model Number48008004S
Device Catalogue Number48008004S
Device Lot Number17708885
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2020
Initial Report FDA Received Date03/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-