• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: I-SENS, INC. RELION PREMIER VOICE BLOOD GLUCOSE TEST SYSTEM; SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

I-SENS, INC. RELION PREMIER VOICE BLOOD GLUCOSE TEST SYSTEM; SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER Back to Search Results
Model Number 780001
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 02/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Meter and test strips involved in incident were returned for evaluation.Initial diagnostic testing indicated a user error finding that the strip release button had been broken.Meter was evaluated with returned strips of specified lot and performed to specification for blood testing.
 
Event or Problem Description
Customer is calling because she feels her readings are higher on the voice meter than her accu-chek meter.Verified she's had the voice meter since (b)(6) 2019.Opened test strips about two weeks ago.Stores supplies in linen closet.Customer is not receiving oxygen therapy.Doesn't have trouble getting a sample of blood.Changes lancets every time she tests.Washes hands with soap and water and allows hands to dry thoroughly.Uses fingertip as a testing site.Seals test strip bottle tightly and immediately after removing a test strip.Customer stated last night she performed a blood test on accu-chek and received 112 and then on the voice and received 167.Customer doesn't have control solution.Explained purpose of the cs test.Customer went into explaining details about when she first noticed that the voice meter was off due to an incident that occurred last week on (b)(6) 2020.She stated on (b)(6) she took a reading at 1:30 am because she woke up due to sweating in her sleep.She received 200 on the voice meter then used her mom's accu-chek and that read 22.She tested about one minute apart.She felt her symptoms did not match the voice meter.She stated the reading was way off.She had a 16-ounce glass of orange juice and three inches sugar in her mouth to increase her blood sugar.She stated that she was shaking and sweating.She didn't have any food or medication two hours prior.She did not contact the paramedics.Her mom treated her with a glucagon.It helped to raise her sugar levels.She went back to sleep.She woke up the next morning and she was 144.Products have not been exposed to any extreme temperatures or humidity.Replaced meter, strips and sent rl.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELION PREMIER VOICE BLOOD GLUCOSE TEST SYSTEM
Common Device Name
SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER
Manufacturer (Section D)
I-SENS, INC.
43, banpo-daero 28-gil
seocho-gu
seoul 06646,
KS 
MDR Report Key9786433
Report Number1832816-2020-00007
Device Sequence Number16768236
Product Code NBW
UDI-Device Identifier00681131169776
UDI-Public(01)00681131169851,(01)00681131169776
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/11/2020
Report Date (Section F) 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/14/2021
Device Model Number780001
Device Catalogue Number780001
Device Lot NumberRY15BBB6J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2020
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date02/11/2020
Device Age7 MO
Event Location Home
Date Report to Manufacturer03/04/2020
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age65 YR
Patient Weight88
Date Report Sent to FDA03/04/2020
-
-