| Model Number |
11590100 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Code Available (3191)
|
| Date of Event |
02/03/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical product: central line.The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.Although requested, patient demographics was not provided.Patient is a neonate as the event occurred in neonatal intensive care unit.
|
| |
|
Event or Problem Description
|
|
It was reported that the nurse found a puddle of fluid on the floor.The tubing was then inspected and found a hole where the leak was coming from.Total parenteral nutrition (tpn) was infusing through a central line at the time of the event.There was a delay in infusion therapy due to changing of iv fluid.The patient was then checked for infection and there was no patient harm but the baby was given prophylactic antibiotics.The event occurred in neonatal intensive care unit (nicu).
|
| |
|
Event or Problem Description
|
|
It was reported that the nurse found a puddle of fluid on the floor.The tubing was then inspected and found a hole where the leak was coming from.Total parenteral nutrition (tpn) was infusing through a central line at the time of the event.There was a delay in infusion therapy due to changing of the iv fluid.The patient was then checked for infection and there was no patient harm, however; the patient was given prophylactic antibiotics.The event occurred in the neonatal intensive care unit (nicu).Although it was confirmed that there was a delay in patient treatment during follow up, it was also noted that there was no patient harm or impact as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report that there was a hole and leak in the tubing was confirmed.The sets were visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.Visual inspection of the set noted the silicone segment was completely separated from the upper fitment.An o-ring indentation was not observed on the separated end of the silicone segment.The ring retainer was not received with the set.No damages (holes) were observed.Functional testing was deemed unnecessary due to the separation and obvious leaking that would occur.Dimensional testing performed found the upper fitment tubing measured to be within the specification(s).The bottom tapered outside diameter of the nipple area measured to be within specification(s).The inside diameter of the silicone segment measured to be within the required specification(s).The root cause of the silicone segment separation was identified as a manufacturing issue for the set¿s ring retainer not assembled onto the set during the assembly process due to cleaning procedures of the equipment and operator error.
|
| |
|