| Model Number |
I.V. STATION |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2020 an i.V.Station device passed an underdosed syringe preparation.The syringe was mishandled after it was weighed and marked acceptable by the device due to improper functionality of the needle remover and robot pincers.The needle remover sensor and needle gripper pincers were replaced and calibrated to prevent recurrence of this issue.The underfilled preparation was identified and there are no known adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
As of january 2020, the establishment registration and listing for this device was updated to omnicell, inc.From aesynt, inc, which was not reflected in the original report submission.
|
| |
|