| Catalog Number |
0375941012 |
| Medical Device Problem Codes |
Break (1069); Particulates (1451)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke and potentially remains in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: a piece of burr remained inside the patient (the side of the burr broke) the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the excessive force applied to the bur head, causing the flutes to rub on the housing surface.Also, a possibility of the bur tip accidentally encountered hard objects during use.Based on the amount of damage observed, it is likely that excessive pressure was maintained for a substantial amount of time and poor suction.The product was returned for investigation and the failure mode will be monitored for future re-occurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke and potentially remains in the patient.
|
| |
|