• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 3.5FR DUAL-LUMEN UVC CATH; CATHETER, UMBILICAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 3.5FR DUAL-LUMEN UVC CATH; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number 8888160531
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.
 
Event or Problem Description
The customer reported that the heparin would not flush through the line when prepping the patient.There was no patient harm.
 
Additional Manufacturer Narrative
H 3: evaluation summary: because the lot number was not available, a manufacturing device history record (dhr) review or product/process changes review for the involved lot number could not be performed.However, all dhrs are reviewed for accuracy prior to product release.One unused catheter was received with its original package opened.The initial test showed that ink did not pass through the secondary extension.Additional testing was performed in order to confirm the location of the occlusion.Based on this testing it was determined the occlusion is at the ¿y¿ junction of the catheter.The occlusion was created by an air pressure chamber not allowing the fluids to enter.A corrective and preventative action (capa) was opened to address this issue.The root cause and corrective or preventive actions will be addressed in this capa.It must be noted that in-process controls are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5FR DUAL-LUMEN UVC CATH
Common Device Name
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
COVIDIEN
edificio b20 calle #2 zona fra
alajuela
CS 
MDR Report Key9787680
Report Number3009211636-2020-00692
Device Sequence Number1866246
Product Code FOS
UDI-Device Identifier20884527005175
UDI-Public20884527005175
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8888160531
Device Catalogue Number8888160531
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/26/2020
Supplement Date Received by Manufacturer02/26/2020
Initial Report FDA Received Date03/04/2020
Supplement Report FDA Received Date05/05/2020
Patient Sequence Number1
-
-