| Model Number |
2 UNIT RED CELL SET W/RC2H FIL |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A blood sample of the employee was taken as well as a sample of the donors' blood for an infectious disease panel to be conducted.Investigation identified evidence that the bowl connector was not completely inserted into the bowl port, and was confirmed per marks presented on blue bowl connector.These marks are produced during sterilization of the disposable and remain permanent until heat is applied to plastic.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, haemonetics was informed by the customer that after the donation an employee was trying to remove the kit and the connector disconnected and sprayed blood in the employee's eye.
|
| |
|