Model Number 1550350-23
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Dyspnea (1816); Fatigue (1849); Low Blood Pressure/ Hypotension (1914); Myocardial Infarction (1969); Thrombosis (2100); Stenosis (2263)
Date of Event 01/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined; however, the reported treatments appear to be related to the operational context of the procedure.The reported patient effects of dyspnea, myocardial infarction, hypotension, thrombosis and stenosis are listed in the xience sierra, everolimus eluting coronary stent system ifu as known patient effects of coronary stenting procedures.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event or Problem Description
Patient id: (b)(6).It was reported that on (b)(6) 2019, a percutaneous transluminal coronary intervention (ptci) was performed on the proximal right coronary artery (rca), 95% stenosed lesion.A 3.25x28mm xience sierra stent was successfully implanted in the proximal rca.On (b)(6) 2020, the patient presented with myocardial infarction, severe in-stent restenosis and in-stent thrombosis.Medications were administered.On (b)(6) 2020 aspiration thrombectomy and revascularization was performed and a xience sierra 3.5x23mm stent was implanted within the xience sierra 3.25x28 stent.The adverse patient event resolved on (b)(6) 2020.(that event was previously reported on another medwatch report number.) on (b)(6) 2020, the patient was admitted to the hospital with fatigue, hypotension, dyspnea and elevated troponin.Myocardial infarction was diagnosed.A heparin drip was initiated.Per imaging, in-stent stenosis was observed in the rca and the left anterior descending (lad) coronary artery had a complete total occlusion.On (b)(6) 2020, another ptci with thrombectomy/thrombolysis was performed in the proximal rca.A non-abbott stent was placed as treatment within the xience sierra stent.Medications were provided and an intra-aorta balloon pump (iabp) was placed.On (b)(6) 2020, another ptci was performed in the mid rca.No additional information was provided regarding this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9789111
Report Number2024168-2020-02022
Device Sequence Number1866610
Product Code NIQ
UDI-Device Identifier08717648227431
UDI-Public08717648227431
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/14/2020
Device Model Number1550350-23
Device Catalogue Number1550350-23
Device Lot Number9080941
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2020
Initial Report FDA Received Date03/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Disability;
Patient Age73 YR
Patient Weight81
-
-