• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 3.5PED
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during use, upon usage on the 7th day, discoloration on the tube end happened.Tube was sterilized and rinsed with alkaline water.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: one sample of device was received for evaluation.The reported event discoloration was confirmed.A visual inspection was performed.It was observed the distal end of the cannula presents a brownish discoloration.The device history record (dhr) was not reviewed since the lot number was not provided.The severity was assessed as 8 (loss of primary function).This product cannot be adequately cleaned and/or sterilized by the user in order to facilitate safe reuse, and is therefore intended for single use.Attempts to clean or sterilize these devices may result in a bio-incompatibility, infection or product event risks to the patient.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez,mx 32590
MX  32590
MDR Report Key9789144
Report Number2936999-2020-00183
Device Sequence Number2411322
Product Code JOH
UDI-Device Identifier30884522026707
UDI-Public30884522026707
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K945513
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3.5PED
Device Catalogue Number3.5PED
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2020
Initial Date Received by Manufacturer 02/08/2020
Supplement Date Received by Manufacturer04/21/2020
Initial Report FDA Received Date03/04/2020
Supplement Report FDA Received Date05/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-