Only the lens was returned, transported inside a lens case in the opened carton.Viscoelastic and reddish colored solution (possibly blood) was observed on the explanted lens.The optic was cut into two pieces, typical in appearance to damage during removal from the eye.The optic portions are adhered in dried solution, one atop the other.One distal haptic tip has areas on the edge that are chipped.The anterior optic surface has an area that is cracked.The center of the optic has cracks that splintered into the material from the cut optic damage.The lens was removed from the lens case to further inspect.There was no foreign material observed other than clinical solution and what appears to be blood.The lens was cleaned separate the optic portions and to remove the solutions to better inspect for the reported "foreign bodies in the iol".Upon cleaning the lens and re-inspecting, there are no foreign bodies observed inside the iol.A qualified viscoelastic was indicated.The root cause cannot be determined for the reported issue.The explanted lens was transported in a multipiece lens case.The lens had damage, viscoelastic, and what appears to be possibly blood.The optic was cut into two pieces, typical in appearance to lens damage during removal from the eye.There was no foreign material observed on the explanted lens surface except for the viscoelastic reddish solution (possibly blood, which would be expected following explant).The lens surfaces were cleaned to inspect for the reported "foreign bodies in the iol".There were no foreign materials observed inside the lens material.The reported complaint was not observed.The manufacturer internal reference number is: (b)(4).
|