|
Product complaint: (b)(4).A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance's related to the reported complaint condition were identified.Summary: twenty-two unopened samples of product were returned for analysis.The complaint sample was not received.During the visual inspection of unopened samples, no defects were found on the packages.The samples were opened, and swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.Per the condition of the samples received, no performance pull off - suture needle was found, and the tested samples met the finished goods requirements.
|
|
It was reported that the patient underwent a spine procedure in 2020 and the suture was used.When trying to use suture, the needle kept popping off the suture, like for a control release suture.Another like suture was used to complete the procedure successfully with no patient consequences.
|