• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT300
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided; best estimate is between (b)(6) 2020 and (b)(6) 2020.(b)(4).Device available for evaluation: yes.Returned to manufacturer on: 2/24/2020 device returned to manufacturer: yes device evaluation: the product was received in the replacement lens¿ daisy wheel along with a completed complaint intake form.Visual inspection with the unaided eye revealed that the lens was received cut in half, which is consistent with a lens that was handled during explant.Additionally, viscoelastic residue was observed on the optic body and haptics.Based on the condition of the return lens no product evaluation could be performed.No product deficiency was identified.The complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxt300 intraocular lens (iol) was implanted in the patient's right (od) eye on (b)(6) 2020.It was later explanted on (b)(6) 2020 due to diplopia anisometropia issues and replaced with a zxr00 iol.An incision enlargement was required for the explant, and no patient outcome was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9790512
Report Number9614546-2020-00101
Device Sequence Number14025506
Product Code POE
UDI-Device Identifier05050474606142
UDI-Public(01)05050474606142(17)230704
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/04/2023
Device Model NumberZXT300
Device Catalogue NumberZXT300U080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/05/2020
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date03/05/2020
Supplement Report FDA Received Date11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-