• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. DISPOSABLE NERVE STIMULATOR - VARI-STIM® III HAND-HELD; STIMULATOR, NERVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. DISPOSABLE NERVE STIMULATOR - VARI-STIM® III HAND-HELD; STIMULATOR, NERVE Back to Search Results
Model Number 8562010
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare provider (hcp) reported that the device didn't work during procedure.Procedure was completed using a back-up product.Patient was alive-no injury.
 
Additional Manufacturer Narrative
H3: analysis found no fault.The device would light when touching the probe tip to the grounding needle per the instructions for use which indicates it was delivering stimulating current.The device was functionally tested in the other two positions with no issues or intermittent behavior while manipulating the switch and tip within the positions.The device was electrically tested per manufacturing instruction with no out of specification condition identified.H6: fdm 4118, fdr 3233 and fdc 11 no longer apply.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE NERVE STIMULATOR - VARI-STIM® III HAND-HELD
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key9792212
Report Number1045254-2020-00135
Device Sequence Number2412074
Product Code ETN
UDI-Device Identifier00643169341814
UDI-Public00643169341814
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K905035
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/08/2020
Device Model Number8562010
Device Catalogue Number8562010
Device Lot Number0215804073
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/11/2020
Supplement Date Received by Manufacturer03/19/2020
Initial Report FDA Received Date03/05/2020
Supplement Report FDA Received Date04/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-