| Model Number |
EVD35-06-100-120 |
| Medical Device Problem Code |
Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician intended to use an everflex entrust with a 5f sheath during treatment of a 100mm calcified lesion in the patient¿s mid and distal left superficial femoral artery (sfa) <(>&<)> popliteal artery (pa).Moderate vessel calcification reported.Vessel diameter reported as 6mm.No damage noted to packaging, nor issues noted when removing the device from the packaging.Ifu was followed and the device was prepped without issue.The thumbscrew/lock-pin was checked for securement prior to procedure.The device was passed through a previously deployed stent without encountering resistance.The lock-pin was removed once the stent was in the correct location and the physician was ready to deploy.There was no damage noted to the deployment mechanism or device handle.The stent only deployed half-way and stopped.The physician cracked open the handle to manually pull outer catheter an release the remaining portion of the stent.The stent was successfully deployed, and the delivery system was removed without issue.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation the entrust stent delivery system was received within a zippered closure plastic pouch.The handle was received split opened, but all handle components were accounted for.The handle sides were examined, and no unusual wear marks were noted.The pull cable and thumbwheel remained attached to the catheter.A kink just distal of the blue isolation sheath/strain relief was noted.It is not known if the kink happened during the procedure or post-procedure given how the device was packaged for return.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|