| Model Number |
MX9342L |
| Medical Device Problem Codes |
Mechanical Problem (1384); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Codes |
Low Blood Pressure/ Hypotension (1914); Patient Problem/Medical Problem (2688)
|
| Date of Event |
01/02/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received indicating that a smiths medical medex stopcocks broke in half causing a "dangerous drop in blood pressure until a new device was located".The reporter indicated that the patient was a vasopressor dependent.The reporter also indicated that spilt was at the contact point (gray/purple connection point between the two yellow stopcock valves).No further adverse events were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: four samples were received to perform an investigation, three of the samples were unused.It was observed on the used sample that the union between stopcocks was broken.Broken sample was inspected under magnification, an elongation was identified in the broken section on one of the stopcocks and on the other stopcock remaining material was observed.Multiple tests were run on the samples which determined that root cause could not be attributable to manufacturing.The most likely root cause was mishandling during use, as an applied force was required to break the devices.The root cause of the problem was determined to be user interface.This remediation mdr was generated under protocol b10009406, as a result of warning letter (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#:(b)(4).
|
| |
|