• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MBT REV TAPERD PRESSFIT REAMER; KNEE INSTRUMENT : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MBT REV TAPERD PRESSFIT REAMER; KNEE INSTRUMENT : REAMERS Back to Search Results
Model Number 2178-63-104
Medical Device Problem Codes Material Twisted/Bent (2981); Naturally Worn (2988)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The reamer has a stopper that slides over it.The stopper is not sliding well because excessive wear on the reamer.No picture was taken as there is nothing to see but it needs to be replaced as it is not working well.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT REV TAPERD PRESSFIT REAMER
Common Device Name
KNEE INSTRUMENT : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9794170
Report Number1818910-2020-06966
Device Sequence Number14004397
Product Code HTO
UDI-Device Identifier10603295096047
UDI-Public10603295096047
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2178-63-104
Device Catalogue Number217863104
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/21/2020
Supplement Date Received by Manufacturer05/04/2020
Initial Report FDA Received Date03/05/2020
Supplement Report FDA Received Date05/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-