|
Product complaint # (b)(4).Date sent to the fda: 04/07/2020.A manufacturing record evaluation was performed for the finished device le6382 batch number, and no non-conformances were identified.Investigation summary: it was reported breakage suture.Unopened samples of product code j214, lot le6382 were received for analysis.During the visual inspection of thirteen samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were observed.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the representative samples, no performance - breakage suture was found and the tested samples met the finished goods requirements.
|