Model Number D12LT
Device Problems Loose or Intermittent Connection (1371); Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.Date of event is 2020.Event day and event month was not reported.Attempts are being made to retrieve the device.To date, the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.
 
Event Description
It was reported that during an unknown procedure, the device's safety lock failed.There were no patient consequences.No additional information is available at this time.
 
Manufacturer Narrative
(b)(4).Date sent: 4/13/2020.D4: batch # r9573n.Investigation summary: the analysis results found that the d12lt device was returned with no damage in the external components.During the functional testing, the bullet retracted, permitting the exposure of the blade during the advancement through the skin test media and returned to safe position upon penetration.The bullet performed as intended without any anomalies noted.It could not be determined what may have caused the event reported.The event described could not be confirmed as the device performed without any difficulties noted.The reported complaint could not be confirmed.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XCEL DILATING TIP TROCAR 12MM
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key9800224
MDR Text Key201013281
Report Number3005075853-2020-01488
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036001577
UDI-Public10705036001577
Combination Product (y/n)N
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/13/2023
Device Model NumberD12LT
Device Catalogue NumberD12LT
Device Lot NumberT40059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2020
Date Manufacturer Received03/23/2020